Skip to content
OpenDue March 31, 2026

Safety Air Monitoring Replacement Gas Chromatograph (GC) Phase III

DEPT OF DEFENSE

About this opportunity

A Gas Chromatograph (GC) instrument is needed for air monitoring in replacement of MINICAMS instruments. Procure one GC instrument with the following specifications: Detector capable of detecting halogen containing chemical agents and compounds that also do not contain a phosphorous or sulfur element. Detector is an alternate to existing Flame Photometric Detector (FPD) and Flame Ionization Detector (FID) Chromatographic column: MXT-1701 phase Air sampling Quick Change Trap Module with flow rate through the trap of at least 500mL/min External mass flow controller, ability to capture the flow meter readings Oven programmable from ambient to 400? Cycle time from injection to injection of 15 minutes or less Visible and audible alarms for system faults and sample concentration detections Integ...

Finding similar opportunities...