Precision Thermocouple Calibrator-Simulator
DEPT OF DEFENSE
About this opportunity
*** Amendment 0003***The purpose of this update is to post more Questions and Answers (Q and A) for this solicitation. All requirements, terms, and conditions within the original solicitation remain firm and unchanged.Q1: Does the 0.1?V resolution requirement in paragraph 2.3 apply across the entire -11 V to 11 V range? A1: No. The intent of the requirement is that the 0.1?V resolution is only required at the lower ranges where it is required to prove accuracy. It is not required across the entire -11 V to 11 V span. Q2: Is there a relationship or dependency between the voltage accuracy requirement (paragraph 2.7) and the temperature accuracy requirement (paragraph 2.5)? A2: No. The Government treats voltage accuracy and temperature accuracy as two separate and distinct characteristics. Sy...